Correlation of RBC indices and RBC histogram obtained by automated hematology analyser with peripheral smear in the diagnosis of anemia


Original Article

Author Details : Rose Maria Thomas*, Sateesh Chavan S

Volume : 9, Issue : 1, Year : 2024

Article Page : 55-61

https://doi.org/10.18231/j.jdpo.2024.010



Suggest article by email

Get Permission

Abstract

Background and Objectives: The impact of early diagnosis and management of anemia on patient outcome has been tremendous. Although conventional microscopic examination of blood using peripheral smear is unequivocal, the advent of automated hematology analyzer has revolutionized the field of clinical hematology and its practice. This study highlights the efficiency and utility of specific parameters obtained by hematology analyzer and its comparison with peripheral smear in the diagnosis of anemia, also determines the sensitivity of the same.
Materials and Methods: 501 Ethylenediamine tetra acetic acid blood samples collected at a tertiary care teaching hospital in Karnataka during one and half years from February 2021 to August 2022, were run on the Sysmex 6-part automated hematology analyzer and simultaneous peripheral smears were prepared and stained using Leishman stain based on standard operating procedures. The histogram and peripheral smear findings were noted.
Results: Among the 501 cases, there were 269(53.7%) cases of Normocytic Normochromic Anemia in which 82.9% showed normal curve, 145(28.9%) cases of Microcytic Hypochromic Anemia in which 92.4% showed shift to left, 31(6.2%) cases of Dimorphic Anemia in which 61.3% showed normal curve contrary to the expected double peak, 24(4.8%) cases of Macrocytic Anemia in which 83.3% showed shift to right, 23(4.6%) cases of Pancytopenia in which 39.1% showed normal curve and 9 cases of Normocytic Hypochromic Anemia.
Conclusion: The Automated hematology Analyzer provides highly valuable information with limited sample and is an essential tool in diagnosis, classification and management of anemia.
 

Keywords: Anemia, Automated analyser, Histogram


How to cite : Thomas R M, Sateesh Chavan S, Correlation of RBC indices and RBC histogram obtained by automated hematology analyser with peripheral smear in the diagnosis of anemia. IP J Diagn Pathol Oncol 2024;9(1):55-61


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.







Article History

Received : 22-11-2023

Accepted : 27-01-2024


View Article

PDF File   Full Text Article


Copyright permission

Get article permission for commercial use

Downlaod

PDF File   XML File   ePub File


Digital Object Identifier (DOI)

Article DOI

https://doi.org/10.18231/j.jdpo.2024.010


Article Metrics






Article Access statistics

Viewed: 784

PDF Downloaded: 235



Medical Abbreviation List